8930EN_4.fm 12/17/10 2:44 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)
MA07445A008
Rev X
Figure 4-13. Fully Retract the Trigger Mechanism.
Figure 4-14. Confirm "Lesion Complete" on Therapy Screen Display.
14. Position the hand piece for the next needle deployment.
Note: After treating one lesion, fully retract the needles and
then rotate the hand piece to the contralateral side. Then
move distally down the urethra approximately 1 cm.
15. Repeat steps 8 through 14 for a new lesion, until the entire
prostate is treated, or complete such treatment as deemed
appropriate.
16. Press the "Exit" icon button (Figure 4-14) to access the "New
Patient" window on the screen display (Figure 4-15).
Prostiva RF Therapy Procedure
Patient Treatment
Needle-retraction window: solid
white
Retracted position
Deployed position
2010-12 8930 System User Guide English
Medtronic Confidential
CRMRef_R06
79
Exit