Radio Frequency Generator Description
Table 2-8. Caution Messages (continued)
No.
Message
61
Right needle off-
temperature too low
May continue with left
needle
Needle may be near
bladder or blood vessel,
or outside prostate
62
Left needle
off-temperature too high
May continue with right
needle
63
Right needle
off-temperature too high
May continue with left
needle
User Interface
Explanation
Right needle is off due to
low temperature; treatment
may be continued with left
needle
Right needle position is not
optimal, but treatment may
be continued with left
needle
Message stays on for ten
seconds
Alarm is 3 short beeps
with varied tones
Left needle is off due to
high temperature;
treatment may be
continued with right needle
Message stays on for ten
seconds
Alarm is 3 short beeps
with varied tones
Right needle is off due to
high temperature;
treatment may be
continued with left needle
Message stays on for ten
seconds
Alarm is 3 short beeps
with varied tones