Table 7-1. Troubleshooting Checklist
Symptom
Impedance is
less than 80 Ω
Temperature
increase on one
needle is slower
than normal
Urethral
temperature rise
is consistent with
one needle
Red "overview"
occludes
vision/blocks
irrigation channel
"Black" vision on
optics
MA07445A008
Rev X
Probable Cause
Needle may be
Retract and
deployed in
redeploy needles
bladder or through
and shields.
bladder neck
Consider
Needle may be in
shortening
a vessel
needles.
Needle may be
Retract and
touching prostatic
redeploy needles
stone
and shields.
Irrigation may still
Turn off irrigation.
be on
Check impedance.
Needle may be in
the bladder or
blood vessel
Needles may be
Stop power
tracking across the
delivery; retract
urethra
and redeploy
needles and
shields.
Blood clots over
Use a Luer lock
optics tip
syringe to flush via
irrigation channel.
This may remove
larger blood clots
with the hand
piece sheath.
Alternatively,
unlock the rod lens
and slowly retract
from hand piece;
gently slide rod
lens in and out of
hand piece to
dislodge any blood
clots
Full bladder
Empty bladder.
Broken optics
Check optics.
Supplemental Information
Troubleshooting Checklist
Actions
Prevention
Ensure needle
deployment is
1 cm distal from
the bladder neck.
Check ultrasound
for evidence and
location of stones.
Check irrigation
before ablation.
Ensure needle
deployment is
1 cm distal from
the bladder neck.
Apply lateral
pressure before
needle
deployment.
Avoid moving and
repositioning of the
hand piece as
much as possible.
Excessive
movement may
cause more
trauma and
bleeding.
Increase flow rate
of irrigation
Irrigate between
lesions to wash
away blood.
Be sure bladder is
appropriately
drained.
Handle optics
carefully.
2010-12 8930 System User Guide English