Case Management and Modifications
The Module disposable and Smart Syringe contain batteries. Battery icons
display status as green (OK), yellow (low- approximately 20 minutes remaining)
or red (depleted). A warning will appear on the screen with a countdown timer
when the disposable has 5 minutes of battery life remaining. If depleted, Display
will automatically require replacement of the disposable or end case.
During a case, if required the Smart Syringe or Module disposable may be
replaced.
Device Status & Replacement
To Replace Disposable
1. Select New (Syringe) or New (Module).
2. Confirm Yes to replace or No to cancel. The system will begin scanning for
the new disposable.
3. Physically replace the disposable.
4. Prime the disposable DO NOT PULL BATTERY TAB OF DISPOSABLE
UNTIL PRIMING IS COMPLETED. System will continue scanning until a
connection is made.
5. Pull battery tab of new disposable and confirm wireless connection.
Possible Alarm Messages
Alarm Message
Threshold % Alarm and Threshold
Reached Alarm
Low Battery Alarm (Disposables)
No Battery Alarm (Disposables)
Possible Error Messages
Error Message
"The connected Syringe is not
calibrated properly. Please connect
another device to continue."
"The connected DyeVert is not
calibrated properly. Please connect
another device to continue."
"No syringe or DyeVert data
received."
"No syringe or pressure module
data received."
"Lost connection with the Syringe.
Attempting to reconnect."
"Lost connection with the DyeVert.
Attempting to reconnect."
"Lost connection with the DyeVert
EZ. Attempting to reconnect."
"Lost connection with the Pressure
Module. Attempting to reconnect."
User Actions
No action required. This alarm is for
notification purposes only if warning tones
are enabled.
If warning tones are enabled, alarm will
sound for approximately 2 seconds. User
must replace the disposable, end the case
or allow the battery to continue depleting
while continuing the case. Monitoring
capability will continue until the No Battery
Alarm sounds.
User must replace the disposable or
continue case without monitoring capability.
If warning tones are enabled an audible
alert will sound in addition to the warning on
the screen.
User Actions
Replace the Smart Syringe.
Replace the DyeVert Plus or DyeVert Plus
EZ Module disposable
End case and replace the Smart Syringe
and DyeVert Plus or DyeVert Plus EZ
Module disposable
End case and replace the Smart Syringe
and Pressure Module disposable
No action required
No action required
No action required
No action required
Error Message
"Bluetooth Fault Detected. Turn
system OFF and ON: 1. OFF:
Press & hold blue power button >
5 secs. 2. ON: Press & release blue
power button. "
"Software Fault Detected. Turn
system OFF and ON: 1. OFF:
Press & hold blue power button >
5 secs. 2. ON: Press & release blue
power button. "
"Cancelling Connection Attempt"
"The syringe cannot be connected.
Replace the disposable and
continue the case, or end the
case?"
"The syringe cannot be connected.
Replace the disposable to start the
case, or return to main menu?"
"The DyeVert cannot be connected.
Replace the disposable and
continue the case, or end the
case?"
"The DyeVert cannot be connected.
Replace the disposable to start the
case, or return to main menu?"
"The DyeVert EZ cannot be
connected. Replace the disposable
and continue the case, or end the
case?"
"The DyeVert EZ cannot be
connected. Replace the disposable
to start the case, or return to main
menu?"
"The Pressure Module cannot be
connected. Replace the disposable
and continue the case, or end the
case?"
"The Pressure Module cannot be
connected. Replace the disposable
to start the case, or return to main
menu?"
"There is no remaining battery
power for the syringe."
"Syringe Battery Level Critical
Replace the Syringe within <timer>"
"There is no remaining battery
power for the DyeVert."
"There is no remaining battery
power for the DyeVert EZ."
"DyeVert Battery Level Critical
Replace the DyeVert within
<timer>"
There is no remaining battery
power for the Pressure Module."
"Pressure Module Battery Level
Critical. Replace the Pressure
Module within <timer>"
"Would you like to continue the
current case started on <date> at
<time>?"
"Unable to connect to the Syringe.
Please replace the unit."
"Unable to connect to the DyeVert.
Please replace the unit."
"Unable to connect to the Pressure
Module. Please replace the unit."
"Please wait as we reconnect to
previously used devices."
"Please wait as we dispose of
previously used devices."
"Database Error: failed to initialize
database."
"Database Error: failed to add
case."
"Database Error: failed to update
case."
"Database Error: failed to mark all
unfinished cases as abandoned."
DyeVert Plus or DyeVert Plus EZ Contrast Reduction System and Contrast
Monitoring System FCC Information
Display Contains FCC ID: Z64-WL18SBMOD
DyeVert Plus and DyeVert Plus EZ Module FCC ID: 2AHUPDV
User Actions
Perform a hard shutdown of the Display by
holding the power button down for >5
seconds, then restart the Display by
pressing the power button once. If the
problem persists, contact Osprey Medical
Perform a hard shutdown of the Display by
holding the power button down for >5
seconds, then restart the Display by
pressing the power button once. If the
problem persists, contact Osprey Medical
No action required
Replace the Smart Syringe or end case
Replace the Smart Syringe and continue
the case or select main menu and cancel
device use
Replace the DyeVert Plus Module
disposable or end case
Replace the DyeVert Plus Module and
continue the case or select main menu and
cancel device use
Replace the DyeVert Plus EZ Module
disposable or end case
Replace the DyeVert Plus EZ Module and
continue the case or select main menu and
cancel device use
Replace the Pressure Module disposable or
end case
Replace the DyeTect Pressure Module and
continue the case or select main menu and
cancel device use
Replace the Smart Syringe
Replace the Smart Syringe within time
remaining on displayed countdown timer.
Replace the DyeVert Plus Module
disposable.
Replace the DyeVert Plus EZ Module
disposable.
Replace the DyeVert Plus or DyeVert Plus
EZ Module disposable within time
remaining on displayed countdown timer.
Replace the DyeTect Pressure Module
disposable.
Replace the DyeTect Pressure Module
disposable within time remaining on
displayed countdown timer.
Select "Yes" to restart the case, "No" to end
the case
Replace the Smart Syringe
Replace the DyeVert Plus or DyeVert Plus
EZ Module disposable
Replace the DyeTect Pressure Module
disposable.
No action required
No action required
Perform a hard shutdown of the Display by
holding the power button down for >5
seconds, then restart the Display by
pressing the power button once. If the
problem persists, contact Osprey Medical
Contact Osprey Medical
Contact Osprey Medical
Contact Osprey Medical
8161-N Jan 2020