Runnin a test
22

RUNNING A TEST

> PATIENT TEST
It is essential that the on-screen prompts are followed throughout the testing process.

5.1 Patient test

Select a new test by pressing the test icon on the home menu. Enter a Patient ID
(1-20 characters) using the onscreen keyboard or a barcode scanner (see Section 9
for more information).
The Actim 1ngeni instrument displays the test number and the patient ID and asks
the user to insert the test cartridge. For sampling and activating the test, refer
to the Actim 1ngeni test instructions for use document. The test begins when
cartridge insertion is detected. Selecting back cancels the test and no data is saved.
The test type and user ID are saved.
The test screen will display:
• Lot number for the test
• Expiry date for the test
• Test number
• Patient ID
• Test type
• User ID
• Test time and date
Remaining time of incubation
An expired test cannot run on the instrument.
The test result will either be positive or negative (see Section 11 for error
messages). The cartridge can be detached from the instrument, when the result
window is visible.