Admin user settings

4.1 Export

The "Export" settings screen allows an admin user to export instrument data
onto an attached USB key.
"Export .csv File" exports a .csv file containing a summary of all of the test
results stored on the instrument. These results are viewable in Excel.
"Export .TSV File" exports a .tsv file containing a summary of all of the test
results stored on the instrument. These results are viewable in Excel.
The user will be given the option to delete all of the test results after export has been performed. The exported
results on the USB stick should be renamed before exporting new results.
"Export Test Diagnostics" enables the admin user to export data where each field value is separated from the
next by a "tab" stop character (e.g. '2 56'). In this case, 2 and 56 will be placed in separate columns.
The user may wish to "Export Instrument Log" and provide this file to the manufacturer during service,
maintenance and repair circumstances. The default setting for this function is "off".
The Actim 1ngeni instrument holds up to 999 results. A warning will appear before the memory becomes full.
The test and QC results must be exported before the memory becomes full, or memory will be overwritten.

4.2 Save Diagnostics

Turning on Save Diagnostics greatly increases the memory usage of the
instrument, as it collects information in the JSON file for troubleshooting
purposes. Only use as directed by instrument support personnel. The default
setting is off.

4.3 Read Only

In the Read Only test mode, the test reaction occurs outside of the instrument.
The user is responsible for the timing and start of the analysis. The Read Only
option is visible in the test options list, if turned on by the admin user. All tests
run in Read Only mode are marked in the test results. The Read Only mode is
useful when tests are read in series.