Herunterladen Inhalt Inhalt Diese Seite drucken

Actim 1ngeni Bedienungsanleitung Seite 37

Vorschau ausblenden Andere Handbücher für 1ngeni:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 152
In-Test Error Dialogs
Focus Out Of Range
The focus of the camera is outside of
limits.
Error: 49
This error will only occur if the test has
the 'Run Focus Check' option enabled
in kinetic.
Error: 50
Analyser Exception
Error: 52
No Image Acquired
Could Not Make Decision, Unknown
Exception
Error: 64
The decision module in the test type
encounters an error.
Could Not Make Decision, Audio Tone
Not Set
This is most likely due to an error in
Error: 65
the logic of the decision algorithm
definition in the test package. The
Audio Tone is not being set by the
algorithm and is required.
Could Not Make Decision, Detailed
Message Not Set
This is most likely due to an error in
Error: 66
the logic of the decision algorithm
definition in the test package.
The required 'Detailed Message' Fields
are not set by the algorithm.
Could Not Make Decision, Icon Not Set
This is most likely due to an error in
the logic of the decision algorithm
Error: 67
definition in the test package.
The required 'Icon' fields are not set
by the algorithm. (RESULT_DECISION_
ICON_X)
Re-power the instrument.
Attempt to run a self test.
Supplier to acquire image and ensure that the internal
validation marks are clearly visible and not obscured by
debris or other materials.
A general analysis error.
Re-power the instrument.
Attempt to run a self test.
If error continues contact supplier.
"Acquire image" of the instrument internals and check the
image quality
is adequate
Camera failed to return an image.
Re-power the instrument.
Attempt to run a self test.
If error continues contact supplier.
"Acquire image" of the instrument internals and check the
image quality
is adequate.
This is most likely due to an error in the logic of the decision
algorithm definition in the test package.
Other Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the audio tone is set correctly by the algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the Detailed Message fields are set correctly by the
algorithm.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the result Icons are set correctly by the algorithm.
Troubleshooting
37

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Actim 1ngeni

Diese Anleitung auch für:

19101ac

Inhaltsverzeichnis