Troubleshooting
40
In-Test Error Dialogs
Error: 86
Normalisation Exception
Could Not Complete Exposure
Error: 96
Calibration, Not Scale Calibration
Strip Analyser Scale Rotate Factor Out
Error: 145
Of Range
Could Not Locate Cartridge, No Edges
Error: 200
Could not locate the cartridge window
in the expected location.
Error: 201
Could Not Locate Cartridge, No Label
Error: 300
Image Level Out Of Range
Error: 301
Voltage Out Of Range
Error: 302
Temperature Out Of Range
Error: 310
Consumable Removed
Error: 320
Barcode Not Found
Error: 321
Barcode Has Expired
Barcode Does Not Match Selected
Error: 322
Test
The normalisation failed to complete.
Re-attempt normalisation.
Contact the Manufacturer.
Could not complete exposure calibration, the exposure
calibration algorithm is not able to complete the primary
scale calibration.
Check that the strip is in the cartridge/strip carrier.
Check that the cartridge has fiducials markings that are
visible and not damaged. Check for contaminants.
Attempt to re-run test.
Could not analyse, the image scale and rotation is out of
range. The instrument requires re-calibration
Check that cartridge is inserted correctly into the slot.
Check for damage or contaminants on cartridge.
Attempt to re-run test
Could not locate the cartridge printed label in the correct
locations.
Check that cartridge is inserted correctly into the slot.
Check for damage or contaminants on cartridge.
Attempt to re-run test.
The camera returned a dark or light image that is out of
range.
Re-power the instrument.
Attempt to run a self test. If error continues contact
supplier.
"Acquire image" of the instrument internals and check the
image quality is adequate
Per-test self test failed.
The system voltages are out of range.
Re-power the instrument.
Attempt to run a self test.
If error continues contact supplier.
Ensure correct power supply is being used
Per-test self test failed.
The instrument temperature is out of range.
Re-power the instrument.
Attempt to run a self test.
The ambient air temperature is out of the acceptable
range.
If error continues contact supplier.
The consumable was removed during image acquisition.
The barcode could not be found in the image.
Ensure that the correct cartridge is being used.
Check that cartridge is inserted correctly into the slot.
Check for damage or contaminants on cartridge
Attempt to re-run test.
The barcode on the cartridge has expired.
Discard cartridge
The barcode on the cartridge does not match the selected
test type. Attempt to re-run test with the correct test type
selected.