Troubleshooting
46
Errors Dialogs
Printer not found!
0569
Please check printer connection and try
again.
Instrument is not normalised!
0570
Testing is locked out.
Please refer to user manual.
Maximum of 99 users reached!
0572
Please delete an existing user ID before
adding a new user.
0591
Cannot proceed!
Invalid Password.
Passwords must be between 8
0593
and 64 characters in length
Return to user default request failed!
0594
Please try again.
0595
Port Number Invalid.
Could not obtain IP Address.
Please check network connection and try
0597
again.
An error has occurred!
0600
Changes were not saved.
0765
Invalid barcode found.
Information Dialog
0256
User deletion completed successfully!
0257
Lot data deleted successfully.
0258
{0} users imported successfully!
Exported {0} users successfully to USB
0259
instrument!
If the user attempts to print a test report before setting
up the instrument printer connections, this error will be
displayed.
a) Try connecting the USB report printer to the instrument.
If the printer is not found after connecting, try to reboot
the instrument.
If the instrument is not successfully normalised prior to
testing, the test results may be incorrect. This instrument
requires normalisation to be complete before the Actim
1ngeni instrument can perform testing.
The instrument is notifying the user that the user list is full.
No more users can be added until some are cleared from
the instrument memory. Please delete a user to be able to
add a new user.
User is unable to proceed with the test selection category
as there are no test types available.
If this occurs contact supplier as the test category may not
be part of the supplied test type package file.
An incorrect or invalid password has been entered. Please
check what was entered then try again.
The instrument has not successfully been returned to user
default.
Re-attempt the request.
If request continues to fail, run a self test.
Contact supplier.
Please enter a valid Port number. An incorrect or invalid
port number has been entered. Please check what was
entered then try again.
The IP Address was unable to be obtained from the
network. Ensure the instrument is connected to a valid
network and try again.
An error occurred writing the changes to the instrument
SD card.
Reboot instrument and try again.
If error continues to occur, run a self test.
Contact supplier.
The barcode scanned is not recognized by the instrument.
Please try again or use the keyboard to enter the details.
The delete user function has completed successfully.
The delete Lot data function has completed successfully.
User import function has completed successfully.
User export function has completed successfully.