In-Test Error Dialogs
Could Not Make Decision, Ratio Title
Not Set
This is most likely due to an error in
Error: 74
the logic of the decision algorithm
definition in the test package.
The required 'Ratio Title' field is not
set by the algorithm.
Could Not Make Decision, Ratio
Output Not Set
This is most likely due to an error in
Error: 75
the logic of the decision algorithm
definition in the test package.
The required 'Ratio Output' fields are
not set by the algorithm
Could Not Make Decision, QC Result
Not Set Correctly
This is most likely due to an error in
Error: 76
the logic of the decision algorithm
definition in the test package.
The required 'QC Result' output is not
set by the algorithm correctly.
Could Not Make Decision, Quant Title
Not Set
This is most likely due to an error in
Error: 80
the logic of the decision algorithm
definition in the test package.
The required 'Quantitative Title'
output fields are not set correctly.
Could Not Make Decision, Quant
Output Not Set
This is most likely due to an error in
Error: 81
the logic of the decision algorithm
definition in the test package.
The required 'Quantitative Output'
fields are not set correctly.
Strip Analyser Could Not Locate
Control Line, No Peak Candidates
Error: 82
The strip analyser could not locate any
trace of a control line in the expected
control line region.
Strip Analyser Could Not Locate Test
Line, No Peak Candidates
Error: 83
The strip analyser could not locate any
trace of a test line in the expected test
line region.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the required number of ratio output fields are defined
correctly.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the required ratio outputs are set correctly.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the QC result flag is set correctly.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the quantitative output fields are set correctly.
Possible causes are:
The test package is not correctly installed – try importing
again.
An incorrect test package is being used with the software
version installed on the instrument. Check test package
release details.
Contact the supplier.
The Supplier will review the test types package to ensure
that the quantitative output fields are set correctly.
Check the control line.
Check for contaminants on the strip.
Attempt to rerun test
If the error continues, contact the supplier.
The Supplier will review the test types package to ensure
that the control line search region is sufficiently wide
enough to enclose the control line.
Attempt to rerun test
If the error continues, contact the supplier.
The Supplier will review the test types package, ensure that
the test line search region is sufficiently wide enough to
enclose the test lines.
Troubleshooting