ADMIN USER SETTINGS
14

4.1 Export

The "Export" settings screen allows an admin user to export instrument data
onto an attached USB key.
"Export Test Results" exports all JSON files.
"Export Summary File" exports a .csv or a .tsv file containing a summary of all
of the test results stored on the instrument. The default setting is .csv. These
results are viewable in Excel.
The user will be given the option to delete all of the test results after export has been performed. The exported
results on the USB stick should be renamed before exporting new results.
The user may wish to "Export Instrument Log" and provide this file to the manufacturer during service,
maintenance and repair circumstances. The default setting for this function is "off".
The Actim 1ngeni instrument holds up to 999 results. A warning will appear before the memory becomes full.
The test and QC results must be exported before the memory becomes full, or memory will be overwritten.

4.2 HTTP Server

HTTP Server mode can be enabled or disabled by the admin user. If configured
to 'On', the results can be viewed via a web browser. The results available are
identical to what is exported to a USB using the instrument's 'Export' function.
To view results navigate to http://<deviceIP>/cgi-bin/testresults.cgi using a web
browser on the same network. For more information refer to section 7.6 Network
access to test results.

4.3 Save Diagnostics

Turning on Save Diagnostics greatly increases the memory usage of the
instrument, as it collects information in the JSON file for troubleshooting
purposes. Only use as directed by instrument support personnel. The default
setting is off.

4.4 Test Mode

In the Read Only test mode, the test reaction occurs outside of the instrument.
The user is responsible for the timing and start of the analysis. The Read Only
option is visible in the test options list, if turned on by the admin user. All tests run
in Read Only mode are marked in the test results. The Read Only mode is useful
when tests are read in series.