• Do not use the siderails (XPS option) as a patient transfer device or surface (for example, to slide a patient from the cot
to another surface).
• Do not position patients with full weight on the siderails (XPS option).
• Do not use the siderails (XPS option) as a push/pull device or to steer the product.
• Do not entangle the restraint straps in the base frame when you raise or lower the cot.
• Always secure the defibrillator platform to the product when you use the defibrillator platform.
• Always use and adjust the straps that are provided with the defibrillator platform to secure the defibrillator.
• Always change the installation location or adjust the straps for your specific defibrillator size or shape.
• Do not load the defibrillator platform above the safe working load of 30 lb (13.6 kg).
• Do not load the equipment hook above the safe working load of 35 lb (15.8 kg).
• Always remove all accessories or equipment from the equipment hook when in the vehicle.
• Do not load the IV pole above the safe working load of 25 lb (11.3 kg).
• Do not load the oxygen bottle holder above the safe working load of 15 lb (6.8 kg).
• Do not use two oxygen bottle holders at the same time.
• Do not load the base storage net above the safe working load of 20 lb (9 kg).
• Always be careful when you retract the base to avoid damaging items that are stored in the base storage net.
• Do not load the backrest storage pouch above the safe working load of 20 lb (9 kg).
• Do not allow the storage pouch to interfere with the operation of the retractable head section.
• Do not load the head end storage flat above the safe working load of 40 lb (18 kg).
• Do not store items under the mattress. Items under the mattress may interfere with product operation.
• Do not steam clean or ultrasonically clean the product.
• Do not exceed 180 °F (82 °C) as the maximum water temperature.
• Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you use a hand held wand to wash the product,
keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
• Always allow to air dry.
• Always remove the battery before you wash the cot.
• Do not clean, service, or perform maintenance while the product is in use.
• Always use authorized parts to avoid the risk of product damage.
6086-209-005 Rev I.1