I I t t e e m m
Base
Wheels
Retractable head section oxygen bottle holder (optional)
Defibrillator platform (optional)
F F o o o o t t e e n n d d f f a a s s t t e e n n e e r r p p a a r r t t r r e e p p l l a a c c e e m m e e n n t t s s c c h h e e d d u u l l e e
For P P e e r r f f o o r r m m a a n n c c e e - - L L O O A A D D compatible cots, you must replace the foot end fastener parts every 18,078 calls. This is to make
sure that the P P e e r r f f o o r r m m a a n n c c e e - - L L O O A A D D remains functional. Follow this call volume time table to remain compliant with this
requirement.
C C a a l l l l s s p p e e r r d d a a y y
6086-209-005 Rev I.1
≤ 7
8
9
10
I I n n s s p p e e c c t t
Fowler bumper and screws
All welds are intact, not cracked, or broken
Check and adjust wheel locks
Straps and clips for wear
Straps not frayed or torn
Latch hooks are intact and secure
Not applicable
M M o o n n t t h h s s
77
67
59
47