6. Commissioning, operation
The following screen will appear:
EN
Fig. 75 - Store report request
Press [YES] to store the report setup and then press again yes to run the report.
Fig. 76 - Report run request
The main screen will be displayed as per below:
Fig. 77 - Report running display
On the bottom menu, the report name (rpt1), the calibration point (1 of 20) and the related reference value (0.00000) will be
shown. Press [STORE] to store the first calibration point: the switch to the second point is done automatically (pnt 2 will be
shown), and so on for all the predefined calibration points. Pay attention to the bargraph in the DUT display: check if point is
inside or outside the limit specified during the DUT assignment.
Once the calibration cycle is defined, the generation of an electrical signal is done automatically.
For the pressure, instead, the operator has to reach the required pressure values using the manual pump and the fine
adjustment.