3. Specifications
Input
Measuring range
signals
10)
Type L
-200 ... 0 °C
EN
0 ... 760 °C
10)
Type C
0 ... 2,000 °C
0 ... 1,300 °C
Type N
Type E
-200 ... 0 °C
0 ... 1,000 °C
10) Precision and accuracy of the e.m.f. values
For measurements with internal cold junction compensation: cold junction error = 0.15 °C
Maximum input voltage: DC ±100 V
Input Impedance: > 100 MΩ
Temperature effect: 0.001 % of reading * |t – t
One year specifications

3.8 Thermocouple simulation

Output
Measuring range
signals
11)
Type J
-190 ... 0 °C
0 ... 1,200 °C
Type K
11)
-160 ... 0 °C
0 ... 1,260 °C
11)
Type T
-130 ... 0 °C
0 ... 400 °C
11)
Type F
0 ... 400 °C
Type R
160 ... 1,760 °C (320 ... 3,200 °F)
Type S
170 ... 1,760 °C (338 ... 3,200 °F)
11)
Type B
920 ... 1,820 °C (1,688 ... 3,308 °F) 0.1 °C
11)
Type U
-160 ... 0 °C
0 ... 400 °C
11)
Type L
-200 ... 0 °C
0 ... 760 °C
Type N
0 ... 1,300 °C
Type E
-200 ... 0 °C
0 ... 1,000 °C
11)
0 ... 2,000 °C
Type C
11) Precision and accuracy of the e.m.f. generation
For temperature simulation with internal cold junction compensation: cold junction error = 0.15 °C

3.9 Environmental parameters module

Parameter
Temperature
Barometric pressure
Relative humidity
14
Linear error
(-328 ... +32 °F)
0.03 °C (0.05 °F) 0.01 °C (0.02 °F) 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS
(32 ... 1,400 °F)
0.04 °C (0.07 °F)
(32 ... 3,632 °F)
0.05 °C (0.09 °F) 0.1 °C
(32 ... 2,372 °F)
0.04 °C (0.07 °F) 0.01 °C (0.02 °F) 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS
(-328 ... +32 °F)
0.03 °C (0.05 °F) 0.01 °C (0.02 °F) 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS
(32 ... 1,832 °F)
0.04 °C (0.07 °F)
| for t : -10 °C ≤ t ≤ 19 °C and 23 °C ≤ t ≤ 50 °C and t
c
14 °F ≤ t ≤ 66.2 °F and 73.4 °F ≤ t ≤ 122 °F and t
Linear error
(-310 ... +32 °F)
0.05 °C (0.09 °F) 0.01 °C (0.02 °F) 0.01 % ±0.003 % FS
(32 ... 2,192 °F)
0.04 °C (0.07 °F)
(-256 ... +32 °F)
0.06 °C (0.11 °F) 0.01 °C (0.02 °F) 0.01 % ±0.003 % FS
(32 ... 2,300 °F)
0.04 °C (0.07 °F)
(-202 ... +32 °F)
0.05 °C (0.09 °F) 0.01 °C (0.02 °F) 0.01 % ±0.003 % FS
(32 ... 752 °F)
0.04 °C (0.07 °F)
(32 ... 752 °F)
0.05 °C (0.09 °F) 0.1 °C
0.04 °C (0.07 °F) 0.1 °C
0.04 °C (0.07 °F) 0.1 °C
(-256 ... +32 °F)
0.04 °C (0.07 °F) 0.01 °C (0.02 °F) 0.01 % ±0.003 % FS
(32 ... 752 °F)
(-328 ... +32 °F)
0.03 °C (0.05 °F) 0.01 °C (0.02 °F) 0.01 % ±0.003 % FS
(32 ... 1,400 °F)
0.04 °C (0.07 °F)
(32 ... 2,372 °F)
0.04 °C (0.07 °F) 0.01 °C (0.02 °F) 0.01 % ±0.003 % FS
(-328 ... +32 °F)
0.03 °C (0.05 °F) 0.01 °C (0.02 °F) 0.01 % ±0.003 % FS
(32 ... 1,832 °F)
0.04 °C (0.07 °F)
(32 ... 3,632 °F)
0.05 °C (0.09 °F) 0.1 °C
Measuring range
-10 ... +50 °C
(14 ... 122 °F)
650 ... 1,150 mbar
(9.43 ...16.68 psi)
10 ... 90 % r. h.
Resolution
Precision
% of rdg ±% FS
(0.18 °F) 0.008 % ±0.002 % FS 0.01 % ±0.003 % FS
= 20 °C
c
= 68 °F
c
Resolution
Precision
% of rdg ±% FS
(0.18 °F) 0.01 % ±0.003 % FS
(0.18 °F) 0.01 % ±0.003 % FS
(0.18 °F) 0.01 % ±0.003 % FS
(0.18 °F) 0.1 °C
(0.18 °F) 0.01 % ±0.003 % FS
(0.18 °F) 0.01 % ±0.003 % FS
Precision
Accuracy
2.7 °C
(4.86 °F)
3.0 °C
4 % FS
5 % FS
12 %
15 %
WIKA operating instructions, model Pascal ET
Accuracy
% of rdg ±% FS
Accuracy
% of rdg ±% FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
0.015 % ±0.003 % FS
Max. resolution
(5.4 °F)
0.1 °C
(0.18 °F)
1 mbar
(0.015 psi)
1 %

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis