10.2.4.1 Calibration presets

The default settings for calibration are stored permanently in this device.
Changing these settings can only be done by Ecolab.
In the calibration presets, the calibration routines are released, which should be
accessible for the respective calibration. Non-approved calibration routines are not visible
in the calibration menu. More calibration presets are explained in the following table.
Parameter
Calibration
rel. Cell constants:
Calibrate
temp coeff.:
Only if calibrate relative cell constant = yes
Ref conductivity.:
Only with calf. Calibrate temp coeff.:
Temp
compensation:
Ref.- temperature: -50 to +150°C
Working
temperature:
Setting option
Yes/No
Yes/No
0 to 9999 µS/cm
* No selection
or analogue inputs /
*IN4 temperature input 1
*IN5 temperature input 2
...
...
IN11 Universal input 2
uncomp.
* IN11 Universal input. 2
comp.
IN12 Universal input. 3
uncomp.
* IN12 Universal input. 3
comp.
* = Please select only the highlighted points
-50 to +150°C
119
Explanation
Allow calibration of the relative cell
constant?
Allow calibration of the temperature
coefficient?
Preset reference conductivity
Temperature input for automatic
temperature measurement at the
measuring point during calibration.
The conductivity of a solution at the
reference temperature and operating
temperature is read during the
calibration process. This results in two
pairs of values (temperature /
conductivity). These pairs of values form
the basis for the calculation of the
temperature coefficient.
The working temperature must differ by
at least 5°C from the reference
temperature.
Device menu
Default
Yes
No
200 µS/cm
No selection
25°C
50 °C
417102269 Rev. 2-08.2020