Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.4

Calibration

The main menu for adjusting the Multronic can be accessed through operating the ↑-key
in the measuring mask.
Main menu
The calibration function is selected through placing the cursor onto the selected item and
operating the ENTER key.
Calibration
module selection
(not available at
Multronic OC!)
The electrodeless C- measuring module that is to be calibrated can now be selected. To do
so, move the cursor onto the corresponding selected item and operate the ENTER key.
Calibration
electrodeless C
The selected measuring range must first of all be checked and possibly changed.
Value range: 0..2 mS / 0..20 mS / 0..200 mS / 0..2000 mS
NOTE
ATTENTION
The continue function cancels the calibration and opens the calibration module selection
menu.
The calibration start function starts the electrodeless C calibration.
The continue function cancels the calibration and opens the calibration module selection
menu.
417101800_Anhang01_LF_induktiv.doc
Conductivity Measuring Module (electrodeless)
It must always be ensured that the correct measuring range has been selected, as
failure to do so can lead to erroneous measurements (overflow if the measuring
range is too small) or inaccuracies (measuring range too large).
Increasing error is to be expected in the case of measurements below 0.5 mS/cm.
- 50 -
Rev. 06-07/10