Analysis Inputs For Calibration Timer - ECOLAB Versatronic Betriebsanleitung

Modulares mehrkanalmessgerät
Vorschau ausblenden Andere Handbücher für Versatronic:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Device menu
Configuration
point
Alarm type
Alarm / event list
Alarm text
Threshold
Hysteresis
Window width
Alarm delay on
Alarm delay off
(
2
) The unit displayed depends on the execution of the plug-in card (type of measurement) in the related analysis slot.

10.4.3.7 Analysis inputs for calibration timer

Analysis inputs: IN7, IN8, IN9, IN10
Go to: Device menu è Configuration è Analogue inputs è Analysis input 1 to 4 è
Configuration timer
Configuration item
Function
Alarm / event list
Alarm text
Calibration interval
417102269 Rev. 2-08.2020
Selection / option
Explanation
Inactive
Choice of alarm type
Min. alarm
Max.- alarm
alarm window
Alarm window invert.
OFF
OFF: Alarm is not displayed
Alarm
Alarm: Alarm is displayed in the alarm
Event
bar of the display
Event: Al. is displayed only in the event
list.
Up to 20 characters of
Text which is displayed when an alarm
text
is in the display and/or is on the event
list.
-99999 to +99999
Threshold at which an alarm is
triggered (depending on the type of
alarm)
0 to 99999
Distance to the threshold
Alarm on =
< Thresh. – Hyst. (With minimum
alarm)
> Thresh. + Hyst. (With max. alarm)
Alarm OFF =
> Threshold (with minimum alarm)
< Threshold (with max. alarm)
0 to 99999
Size of the alert window (with alarm
type "alarm window") or of the window
in which no alarm is triggered
( "inverted alarm window" in the alarm
type)
0 to 999 s
Time delay between occurrence of the
alarm condition and the alarm turning
on
0 to 999 s
Time delay between end of the alarm
set to on and the alarm turning off
Selection / option
Inactive active
OFF
Event
Alarm
Text box
0 - 9999
158
Explanation
Enable calibration timer
OFF: No entry
Event: At the end of the
calibration interval, the
text from "Alarm Text" is
written in the event list.
Alarm: At the end of the
calibration interval, the
text "Alarm Text" is written
in the alarm bar. At the
same time, an entry is
added to the event list
Text shown as alarm text
Time after which a
calibration request is
displayed
Default
Inactive
Alarm
Alarm 2
Temp.input 1
0,00
0
0
0 sec
0 sec
Default
Inactive
Event
Request cal. Ch. X
10 days