Configuration
point
Signal tolerance
band w
Window width
(
1
) The unit for the window width comes from the parameter "Signal tolerance band x" in this table.

Timing diagram

Fig. 88: Timing diagram
1
Follow-up time timer
2
Output signal
3
Signal for timer start
4
Lead time Timer start
5
Timer time

10.4.10.2 Time switch

Timer settings for a weekly scheduling program
Requirement: The "Timer function" must be configured as a "timer".
For each day of the week, you can set up to 4 ON and OFF times.
Call switching time settings:
Device menu è Configuration è Timer è Timer 1 to 2 è Zeitschalter è Monday to
Sunday
Configuration item
Timer function:
Switch time 1 to 4
Off times 1 to 4
Selection / option
Select from analogue sector Selection of the analogue signal, for
0 to 99999 (
1
)
Selection /
settings
Explanation

Timer

Weekly timer function:
Time switch
up to 4 adjustable switch-on and off times per week
Time switch with
runtime
hh:mm:ss
If the switch-on time is exceeded, the timer is set to
binary value = 1 (ON).
hh:mm:ss
Once the time for the switch-off is exceeded, the
timer is set to binary value = 0 (off).
169
Explanation
which the signal "tolerance band x"
must not be more than the "window
width" deviate so that the timer can
run
Maximum amount of deviation
| X - w |
6
Follow-up time, timer end
7
Lead time Timer start
8
Timer time
9
Follow-up time, timer end
Device menu
Default
No selection
0.0000
Default
Time switch
00:00:00
00:00:00
417102269 Rev. 2-08.2020