9.1. Pump Failure
An unusual or high-pitched sound indicates that pump failure is occurring. Replace the Affinity CP centrifugal blood pump
immediately.
9.2. Lack of Flow
If the pump does not rotate or there is no flow, consider the following:
Verify that the pump and the drive motor are properly connected.
If using the adapter, consider the following:
Verify that the pump is properly connected to the adapter. Ensure that the pump is secure under the retaining ledge and the
retaining latch of the adapter.
Verify that the adapter is properly connected to the drive motor.
Reduce the pump speed to zero and restart the drive motor.
Follow hospital protocol to reestablish flow.
9.3. Unusual Noise
If the pump vibrates or makes a grinding noise, consider the following:
Consider replacing the pump if the vibration or grinding noise continues.
If using the adapter, consider the following:
Verify that the pump is properly connected to the adapter. Ensure that the pump is secure under the retaining ledge and the
retaining latch of the adapter.
Verify that the adapter is properly connected to the drive motor.
Reduce the pump speed to zero and restart the drive motor.
Follow hospital protocol to reestablish flow.
Consider replacing the pump or the adapter if the vibration or grinding noise continues.
10. Storage and Cleaning of the Affinity™ CP Adapter
Warning: Do not sterilize the adapter. Sterilization may damage the adapter.
Caution: Do not store at extremes of temperature and humidity. Avoid storing the adapter in direct sunlight.
Caution: Do not use alcohol-based solutions, other organic solvents, or concentrated acids or bases on the adapter. The adapter
may be damaged by these solutions.
Follow your hospital's protocol to clean or disinfect the adapter and consider the following:
Use only hospital-approved disinfectants to clean the adapter. The following solutions are acceptable for cleaning or
disinfecting the adapter: water, Betadine™, hydrogen peroxide, and mild detergent. After cleaning, wipe the adapter with a
sponge or soft cloth and water.
Immediately remove any contamination from the adapter by using a sponge or soft cloth and water.
If contamination cannot be removed with a sponge or soft cloth and water, the adapter may be immersed in water for up to
10 minutes. Do not immerse the adapter in any other liquids.
11. Service
Warning: The adapter is not serviceable. Do not disassemble the adapter because the performance of the adapter may be
adversely affected. If the adapter has been tampered with, do not use it.
12. Specifications for the Affinity™ CP Centrifugal Blood Pump with Balance™ Biosurface
Part number:
Priming volume:
Inlet/Outlet ID:
Maximum operating pressure:
Maximum flow rate:
BBAP40
40 ml (approximate)
9.5 mm (3/8 in)
760 mm Hg (101 kPa)
10 L/min