Caution: Appropriate blood anticoagulation protocol should be followed and anticoagulation should be routinely monitored during
all procedures. The benefits of extracorporeal support must be weighed against the risk of systemic anticoagulation and must be
assessed by the prescribing physician.
Caution: Do not store products coated with Balance Biosurface above 50°C (122°F).
8. Instructions for Use
If using the Affinity CP centrifugal blood pump without the Affinity CP adapter, refer to Sections 8.1, 8.2, 8.7, and 8.8 of these
instructions for use.
If using the Affinity CP centrifugal blood pump with the Affinity CP adapter, refer to Sections 8.3, 8.4, 8.5, 8.6, 8.7, and 8.8 of these
instructions for use.
8.1. Installing the Affinity™ CP Centrifugal Blood Pump on the Medtronic External Drive Motor
Model 560A
Connect the external drive motor model 560A to the appropriate Medtronic controller. For complete installation information, refer
to the appropriate Medtronic controller operator's manual.
1. Slide the Affinity CP centrifugal blood pump into the external drive motor model 560A (Figure 3) until the thumb latch pin
engages.
1
Note: Whenever possible, install the pump at the lowest point in the circuit, directly below the reservoir.
2. Prime the centrifugal blood pump according to the procedures described in Section 8.7 of these instructions for use.
Caution: Operating the centrifugal blood pump without priming the pump may cause damage to the centrifugal blood pump.
3. Turn the rpm knob to the clicked-off position. As the knob is turned to zero, the user will feel a slight pressure and hear a click.
The black button on the knob serves as the pointer to the rpm settings.
Warning: Use of the Affinity CP centrifugal blood pump beyond the labeled recommendations may result in failure of the
centrifugal blood pump, reduced pumping capacity, leaks, excessive blood trauma, or degradation or corrosion of blood contact
materials that may pass through the blood to the patient.
Caution: Installing the centrifugal blood pump with the drive motor revolving may harm the pump.
4. Proceed with extracorporeal circulation. Refer to the instructions for use supplied with each device.
8.2. Removing the Affinity™ CP Centrifugal Blood Pump from the Medtronic External Drive Motor
Model 560A
Remove and discard the pump after each procedure. Discard the pump according to local and hospital protocols.
1. Push the drive motor thumb latch to release the centrifugal blood pump (Figure 3).
2. Slide the centrifugal blood pump out of the drive motor receptacle (Figure 3).
8.3. Inserting the Affinity™ CP Adapter
The Medtronic Affinity CP adapter (Figure 4) is intended to be used to enable the Medtronic Affinity CP centrifugal blood pump
(model BBAP40) to be operated with the Stöckert and Sorin centrifugal pump systems or the Sarns and Terumo centrifugal
systems.
14
Instructions for Use
Figure 3. Installing and removing the Affinity CP centrifugal blood pump
English
1. Thumb latch with locking pin on external drive motor
model 560A