Warning: Ensure that the adapter is properly attached to the drive motor. Verify that the adapter is locked onto the drive motor.
Improper attachment may adversely affect adapter performance.
Caution: When inserting the adapter on the drive motor, make sure that the underside of the adapter is in close contact with
the receptacle of the drive motor and that the retaining key is in the locked position.
8.5. Inserting the Affinity™ CP Adapter on the Sarns™ and Terumo™ Centrifugal Systems
Caution: For complete installation information, refer to the instructions for use provided with the Sarns and Terumo centrifugal
systems.
Caution: During use, do not bump or strike the adapter or the circuit components to reduce the risk of the adapter and pump
detaching from the drive motor or any tubing connections. If detachment occurs, refer to Section 9.
Note: Before inserting the adapter on the drive motor, inspect the adapter for visible biocontamination, such as blood or foreign
material. If necessary, clean the adapter or drive motor receptacle according to the instructions in Section 10.
Note: These instructions also apply when the adapter is used with a compatible handcrank from the Sarns and Terumo centrifugal
systems.
Insert the Affinity CP adapter on the drive motor of the Sarns and Terumo centrifugal systems by performing the following steps:
1. Open the receptacle latch of the drive motor and remove the protective cover from the drive motor receptacle.
2. Position the bottom of the adapter over the drive motor receptacle. Slip the adapter under the receptacle clip (Figure 6).
1
4
5
Figure 6. Inserting the Affinity CP adapter into the Sarns and Terumo centrifugal system's drive motor
3. Open the receptacle latch and press the adapter into place. Release the latch to secure the adapter in place.
Warning: Ensure that the adapter is properly attached to the drive motor. Verify that the adapter is locked onto the drive motor.
Improper attachment may adversely affect adapter performance.
Caution: When inserting the adapter on the drive motor, make sure that the underside of the adapter is in close contact with
the receptacle of the drive motor and that the receptacle latch is in the locked position.
8.6. Inserting the Affinity™ CP Centrifugal Blood Pump on the Affinity™ CP Adapter
Insert the Affinity CP centrifugal blood pump on the Affinity CP adapter by performing the following steps:
1. Position the bottom of the pump over the adapter receptacle. Slip the pump under the retaining ledge.
2. Open the retaining latch and press the pump into place (Figure 7). Release the retaining latch and adjust the pump until the
retaining latch locks.
1
2
3
Figure 7. Inserting the Affinity CP centrifugal blood pump onto the Affinity CP adapter
16
Instructions for Use
3
2
English
1. Affinity CP adapter
2. Drive motor receptacle
3. Receptacle clip
4. Receptacle latch
5. Drive motor
1. Affinity CP centrifugal blood pump
2. Affinity CP adapter
3. Drive motor