Herunterladen Inhalt Inhalt Diese Seite drucken

DeVilbiss SmartLink-System Bedienungsanleitung Seite 36

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

EN
Setting Error
Error occurred while updating device settings. This displays until acknowledged or for 15 minutes.
Contact Provider
Card Removed
Data card is removed from the module. This displays for ten seconds.
Send to Provider
Setup Required
Module and CPAP are not matched. The module is unable to record data; however, the CPAP will continue to provide
Contact Provider
therapy. Displays until the module is removed or data is cleared by holding Down Arrow key for 10 seconds.
A module fault has occurred (Clock Error, Memory Error, Setting Error, etc).
E00 Module Fault
Contact Provider
Firmware is being updated. This displays until update is complete and system operation is normal.
Firmware Update
Please Wait . . .