STORING

Storing liquid nitrogen
To ensure a clean supply of liquid nitrogen the storage tank (dewar)
must be completely emptied prior to refilling 3 - 4 times per year. If
necessary, water from icecrystals remaining inside the tank can be dried
out by creating air circulation (e.g. using a vacuum cleaner for approx. 30
minutes).
Storing the CryoPro® device
At the end of the workday it is highly recommended to empty the unit
by removing the lid and discarding any excess liquid nitrogen. Do NOT
pour it back into the dewar to avoid contamination. Once the bottle is
empty it is recommended to store the bottle upside down for a minimum
of 30 minutes before applying the lid. In order to avoid accumulation of
condensation the CryoPro® should be stored with the lid on.
The empty device and accessories must be stored at regular room
temperature and humidity. There are no special storage requirements
and no restrictions on shelf-life. Contact probes do not need to be stored
in sterile condition but must be sterilized before use.
The lifetime of the device and the accessories is 5 years. The lifetime of
the contact probes is 5 years or 100 reprocessing cycles.

MAINTENANCE

No preventive maintenance is needed.

DISPOSAL

CryoPro® and metal accessories can be disposed of as metal scrap
or returned to Cortex Technology for disposal. It is recommended
to decontaminate contact probes before disposal or dispose as
contaminated materials.

PERFORMANCE CHARACTERISTICS

Device
LN2 capacity
CryoPro® MAXI
CryoPro® MINI
The flow rate and the velocity of the LN2 and the maximum treatment time
depends on the spray tip (A-D) mounted:
Spray tip
Hole diameter (mm)
Average flow within
the first minute (mL/min)
Average velocity (m/s)
Estimated total treatment
time in 1 min sessions
(MINI/MAXI) (minutes)
Static holding time
500 ml
20-24 hours
350 ml
12-14 hours
A
B
C
1.00
0.75
0.55
54
28
15
0.75
0.88
0.92
6/9
12/17
22/32
Approximate time from closing the lid after filling the CryoPro®
until it is ready for use
CryoPro® thermal
state when filled
Fill level
CryoPro® Maxi
(500 mL)
CryoPro® Mini
(350 mL)

EXPLANATION OF SYMBOLS

Graphic
D
0.45
9
0.98
36/52
Room temperature (22±1 °C)
Maximum
1/3 of max.
volume
volume
1 minute and
30 seconds
30 seconds
25 seconds
1 minute
Description
Indicates the medical device manufacturer.
Indicates the date when the medical device
was manufactured.
Indicates the manufacturer's catalogue number
so that the medical device can be identified.
Indicates the manufacturer's serial number so
that a specific medical device can be identified.
Indicates the need for user to consult the
instructions for use.
Indicates the item is a medical device.
Indicates a carrier that contains unique device
identifier information.
CE-marking with Notified Body reference no.
xxxx. Signifies European technical conformity.
Caution: Federal (US) law restricts this device
to sale by or on the order of a physician.
Cooled
(refilled)
1/3 of max.
volume
15 minutes
10 minutes
45
Inhaltsverzeichnis
loading

Verwandte Produkte für Cortex CryoPro

Inhaltsverzeichnis