6.4.3
Flushing the Tubular Shaft
Danger of injury from flushing water!
During the flushing of the inner and outer tubes with the flushing adapters provided for this purpose,
there is risk of splashing!
• Perform the flushing process carefully.
• Protect eye area!
To ensure thorough and reliable internal cleaning of the inner and outer tubes, use the respective flushing
adapter provided.
• Use the water jet gun to clean the inner and outer tubes via the Luer lock connection of the
corresponding flushing adapter for at least 5 s in a pulsating manner.
• Clean in ultrasonicator with e.g. neodisher® MediClean forte, working solution 0.5 % at max. 40 °C
(104 °F) for 5 min.
• Flush the inner and outer tubes using the water jet gun for at least 5 s in a pulsating manner.
• Check parts, especially the jaw part, for cleanliness.
• Check for patency of the tubes.
• Repeat the cleaning process if required.
Use of hydrogen peroxide can lead to premature wear of the instrument and thus reduce its lifetime.
78
Bipolar Sealing System marSeal5 plus/ - Maryland
• Plug the silicone hose of the flushing adapter
onto the inner tube.
• Plug the silicone hose of the flushing adapter
onto the outer tube.
• Alternatively, the inner and outer tubes can be
cleaned directly with the water jet gun.
Instructions for Use
Revision 06