5.2

Before First Use and Before Each Further Use

• Observe the safety notes, see section 3.11 "Warnings", page 58
• A functional check must be carried out before each use, see section 7.1 "Functional Check", page 88
Risk of infection in case of non-sterile handling!
Non-sterile handling and improper sterilization can cause severe health risks for the patient!
The instruments must be cleaned and sterilized before the first and any further use, see
section 5.2 "Before First Use and Before Each Further Use", page 68.
• Products in sterile packaging, such as the sterile blade holder REF 80-633-09-04, can be used directly
from their sterile packaging without further pre-treatment.
• After each use, one indicator must be removed from the cycle counter before re-sterilization, see
section 3.10 "Limitations of Use", page 58.
5.3

Application / Intraoperative Procedure

5.3.1
Connection to the Electrosurgical Unit maxium®
The bipolar sealing instrument marSeal5 plus is intended exclusively for use with the electrosurgical unit
maxium® (REF 80-042-00-04, REF 80-042-02-04 and REF 80-042-04-04, type ME 402) from software
version V3.423, or new developments by the manfuacturer. The IQ connector of the marSeal5 plus can be
connected to the bipolar socket B 1 or B 2 of the electrosurgical units maxium®. The current type
SealSafe® IQ is selected automatically. It can be used exclusively with IQ instruments (Plug & Play)!
For correct operation and adjustment, the Instructions for Use of the electrosurgical unit maxium® must
be observed.
• Connect the plug or connection cable (8) of the marSeal5 plus to the bipolar HF output B1 or B2 of the
electrosurgical unit maxium®.
− The electrosurgical unit maxium® automatically supplies the required current type SealSafe® IQ with a factory
default setting G3.
If a foot switch is required for activation of the sealing current, it must be selected separately.
68
Bipolar Sealing System marSeal5 plus/ - Maryland
Instructions for Use
Revision 06