Bipolar Sealing System marSeal5 plus/ - Maryland
Instructions for Use
Possible danger of injury to the patient from residues on the medical device!
Particles from cloths and brushes may get stuck on products with rough surfaces, threads, sharp edges,
narrow gaps or the like.
• Use only soft, clean and lint-free cloths for drying.
• The brushes for ducts must be a little bit bigger than the inner diameter of the duct, and the shaft of
the brush must be at least as long as the duct.
6.4

Cleaning and Disinfection

6.4.1

Manual Cleaning and Disinfection

Recommendations for manual cleaning: Validated cleaning and disinfection procedure acc. to
DIN EN ISO 15883. Deviating cleaning procedures, validated acc. to DIN ISO 15883, are possible.
Manual pre-cleaning serves as preparation for machine cleaning.
6.4.2

Cleaning the Jaw Part

Risk of damaging the jaw part!
The supplied cleaning tool is used to clean the blade guide in the jaw part and the joint area. It must not
be used for scraping off sealing residues on the electrodes, as the polished electrode surface could be
scratched!
• To protect the electrode surface, use soft brushes, damp swabs or paint brushes.
• Clean the blade guide in the jaw part and the joint area using the cleaning tool.
• Clean the jaw part using nylon brushes, swabs and/or paint brushes until the coarse contamination has
been removed.
Fig. 6-1:
Cleaning of the jaw part using the cleaning tool