Bipolar Sealing System marSeal5 plus/ - Maryland
Instructions for Use
Danger of infection and injury from sharp blades!
When the tubular shaft is dismantled, the integrated blade (2) can be pushed through the tubular shaft
and out of the open jaw part by accidentally pressing on the rear end of the blade holder (1).
• Avoid contact with the sharp blade (2)!
The working method must at all times be carried in such a manner that no contamination of the sterile
surgical gloves and the sterile blade holder occurs.
• Remove the sterile blade holder from the sterile
packaging (on the side of the knurled nut).
Revision 06
• Remove protective cap.
1 Outer tube
2 Inner tube
3 Blade Holder
• Slide the blade holder (3) into the inner tube (2)
from the rear.
• If necessary, turn the blade holder slightly in the
last third until it can be pushed into the jaw
part.
87