6.3.4
Disposal of the Blade Holder
Danger of infection and risk of damage in case of improper processing!
Non-sterile application can cause severe health risks for the patient!
• Sterile and non-sterile blade holders are single-use products and must be disposed of after use in
accordance with the applicable regulations!
• Never process and/or reuse blade holders! Processing the blade holder can damage the instrument!
After application and after the product has exceeded its lifetime, we recommend disposing of the blade
holders in a germ-proof and puncture-proof container, see section 6.6 "Packaging", page 81.
6.3.5
Disassembly of the Tubular Shaft into Inner and Outer Tube
• Press the button on the rotary wheel (marked in
white in the illustration) and slide the inner tube
to the front.
The following figures show the inner and the outer tube after disassembly (the shaft length varies
depending on the model)
Inner tube with jaw part
Outer tube
The outer tube and the inner tube are marked with a continuous serial number.
• Ensure that the two tubes stay together during the cleaning procedure and further use.
76
Bipolar Sealing System marSeal5 plus/ - Maryland
• Keep the button on the rotary wheel pressed, pull
out the inner tube (1) from the outer tube (2).
Instructions for Use
Revision 06