Risk of infection in case of non-sterile handling!
Non-sterile use and improper sterilization can cause severe health risks for the patient!
• Sterility of the sterile blade holder (REF 80-633-09-04, REF 80-633-10-04, REF 80-633-15-04) can be
guaranteed only if the packaging is undamaged.
• Do not use after the expiration date indicated on the packaging.
• Insert the sterile-packaged new blade holder only after the instrument has been sterilized and only in
the sterile operating area.
• Blade holders are single-use items.
− Remove the blade holder following each application!
− Dispose of used blade holders!
• Never process blade holders!
• Never reuse blade holders!
Danger of injury or tissue damage from incorrect operation!
• Before starting the sealing process, verify that you have grasped the correct structures, in order to
prevent inadvertent capture of healthy structures such as ureter, nerve tracts, intestinal loops, etc.!
• Permanent occlusion is possible for vessels up to 7 mm in diameter. Placing two seals side by side is
recommended for large vessels. The integrated blade can then be used to cut between these. The
maximum sealing length of the instrument is 20 mm, of which approx. 19 mm are cut.
• Ensure that the tissue structures to be sealed are not too thin. For the sealing process, the amount of
tissue must sufficiently fill the space between the electrodes after latching them!