6.3

Preparation before Cleaning

6.3.1

Disassembly of the Instrument

• Clean the marSeal5 plus in disassembled state.
The instrument must always be cleaned manually and mechanically.
It is recommended to carry out disassembly immediately after the application in the operating theater.
To separate the tubular shaft from the handle, it must not be locked. The jaw part of the tubular shaft
must be open!
6.3.2

Disassembly into Handle, Tubular Shaft and Blade Holder

• To remove the tubular shaft, open the handle
by pushing apart the handle plates.
74
Bipolar Sealing System marSeal5 plus/ - Maryland
Handle disassembled into 2 handle plates
Instructions for Use
Revision 06