6.10.2 Checking the Blade Feed
6.11 Inserting a Sterile Blade Holder
The sterile blade holder can be inserted directly from the sterile packaging into the tubular shaft without
further pre-treatment.
• Do not use the sterile blade holder until after sterilization of the instrument and only in the sterile
operating area.
Danger of infection from damaged sterile packaging!
The sterile blade holder is a single-use product and STERILE on delivery.
A sterile blade holder removed from damaged or opened sterile packaging and from packaging with an
expired use-by date must be considered non-sterile.
• Do not use! The sterility of the blade holder can be ensured only if the packaging is undamaged.
86
Bipolar Sealing System marSeal5 plus/ - Maryland
• Tighten the knurled nut (1).
The blade holder in non-sterile packaging is secured.
• To check the blade feed, press against the rear of
the blade holder (1).
Instructions for Use
Revision 06