6.4.6

Drying

Ensure reliable drying!
Reliable drying is crucial for the success of subsequent sterilization.
6.5

Inspection, Functional Check, Maintenance

6.5.1

Inspection and Functional Check

Further use is confirmed by the successful inspection of the product. The inspection release and packing
into a sterile barrier system releases the product for the next application.
• The products must be macroscopically clean, i.e. free from visible contamination, after each cleaning
process.
• Stained instruments must be discarded immediately and fed into special treatment.
• All moving parts must be inspected with particular care.
• In case of defects or damage, discard the instruments immediately.
• After each cleaning procedure, treat the marSeal5 plus with a physiologically harmless instrument oil or
instrument care spray in the area of the jaw part as well as on the joints and sliding guides of the
handles.
Instrument care agents or surface tension reducing agents may be added as long as they are approved
for the medical instruments and validated in the process.
• After the inspection, store the marSeal5 plus in a specially designed tray basket to avoid transport
damage, and protect it against damage using suitable devices.