Danger of injury for patient and user from improper application!
Improper application and abuse or misuse may result in injury to patient and user and/or premature wear
of the sealing system!
• Clean and sterilize the marSeal5 plus before each use, see section 6 "Cleaning, Disinfection and
Sterilization", page 72.
• Insert and pull out the instrument into/from the trocar only with the jaw part closed!
• Use the bipolar sealing system only for its intended purpose, see section 7.1 "Functional Check",
page 88.
• Avoid touching exposed metal parts on the instrument (handle plate must be properly mounted on the
handle), see section 4.1 "Assembling the Handle", page 61.
• During the functional check, see section 7.1 " Functional Check", page 88, there is a danger of injury
from the sharp blade! The sterile working method must at all times be carried out in such a manner
that no contamination of the sterile surgical gloves and the sterile blade holder occurs.
− Avoid touching the sharp blade!
− Ensure that the blade never touches hard objects.
• Route the connection cable so that it touches neither the patient nor other cables, and does not form
an obstacle. Protect the cable from mechanical damage (running over, crushing, bending)!
• Use of high-frequency devices can interfere with other devices!
Possible danger to life for third parties in case of shipment of contaminated products!
In the case of return shipments, send only clean and disinfected products in sterile packaging.
60
Bipolar Sealing System marSeal5 plus/ - Maryland
Instructions for Use
Revision 06