Bipolar Sealing System marSeal5 plus/ - Maryland
Instructions for Use
6.5.2
Maintenance
"Maintenance" refers to the application of instrument oil or instrument milk (emulsion of white oil in
water). The joints of the instruments must be treated with steam-sterilizable care agents based on paraffin
oil.
LubriPen®, REF 55-997-01-04 for the maintenance of surgical instruments.
6.6
Packaging
Approved sterilization packaging (e.g. conforming to EN 868, ISO 11607) must be used for sterilization,
subsequent transportation and storage.
6.7
Sterilization
6.7.1
Before Sterilization
6.7.1.1
Cycle Counter
A cycle counter with 55 indicators is located on the device-side connector of the instruments. The cycle
counter makes it possible to easily mark the number of applications already completed.
Revision 06
A spare instrument must be ordered well before the
guaranteed applications have been completed, in order to
ensure continuous usability of the bipolar vessel sealing
instrument.
81