Bipolar Sealing System marSeal5 plus/ - Maryland
Instructions for Use
3.10 Limitations of Use
Danger of injury in case of insufficient qualification of the user!
Safe use of the marSeal5 plus in combination with the current type SealSafe® IQ requires the user to be
familiar with the technology and applications of electrosurgery.
Instruments for electrosurgery may be used only by persons who have been specially trained or instructed
in their use!
Danger of injury from improper application and abuse or misuse!
Tube ligation with the marSeal5 plus is not a reliable procedure for contraception!
Not for use on the appendix!
Danger of injury from an excessive number of processing cycles!
With each application, the wear of the instrument progresses. The instrument is designed for maximum of
50 applications, provided the instrument is cleaned and processed according to the information in the
Instructions for Use and subjected only to loads typical for the respective indication. The possible number
of applications depends heavily on the type of application, preparation and maintenance of the
instrument. After 50 applications the instrument must be disposed and replaced with a new instrument.
• Use the cycle counter to mark the processing cycles already performed.
3.11 Warnings
Non-compliance with the safety measures may result in serious or even fatal injury to the patient!
Danger of explosion and fire from flammable gases!
No flammable anesthetic gases or oxidizing gases may be used in combination with the bipolar sealing
system marSeal5 plus.
• Before sealing, ensure that no endogenous gases are present in the area of application.
58