Bipolar Sealing System marSeal5 plus/ - Maryland
Instructions for Use
Danger of injury from an excessive number of processing cycles!
With each application, the wear of the instrument progresses.The manfuacturer guarantees reliable
functioning for a maximum of 50 cycles, provided the instrument is cleaned and processed according to
the information in the Instructions for Use and subjected only to loads typical for the respective
indication. The manufacturer recommends discontinuing use after this time and replacing the
marSeal5 plus to ensure risk-free and safe operation.
• Use the cycle counter to mark the processing cycles already performed.
Product lifetime is usually determined by wear and damage during instrument use. Possible damage
includes: Loss of insulation on the cable, dirt on the jaw part that cannot be removed, damaged or
scratched contact surfaces with tissue adhesion, deterioration of function due to decreasing compressive
force in the jaw part, loosening or wear of components, bent tubular shaft.
• At the end of the product lifetime, dispose of the marSeal5 plus in accordance with national regulations,
see section 9.3, "Disposal", page 92.
6.2

Pre-Treatment at the Application Location prior to Cleaning

It is recommended to carry out disassembly immediately after the application in the operating theater.
To separate the tubular shaft from the handle, it must not be locked. The jaw part of the tubular shaft
must be open!
Conduct the cleaning and disinfecting of contaminated instruments as soon as possible after use.
The following generally applies:
The sooner the inner lumina are cleaned, the better the result will be. Should it not be possible to comply
with this time due to the length of the application or due to organizational aspects, it is the user's own
responsibility to specify and validate measures in order to prevent drying of the contaminants, or to ensure
success of the cleaning despite drying, respectively.
We generally recommend dry return transport in a closed transport container without addition of liquids,
cleaning agents or disinfectants.