Instructions for Use
The following evaluation criteria are an aid to the detection of defects on the product or similar products.
A comprehensive evaluation matrix is available from Gebrüder Martin under REF 90-460-01-04.
5.5.2

Maintenance

"Maintenance" refers to the application of instrument oil or instrument milk (emulsion of white oil in
water). The joints of the instruments must be treated with steam-sterilizable care agents based on paraffin
oil.
LubriPen®, REF 55-997-01-04 by Gebrüder Martin for the maintenance of surgical instruments.
Implants must under no circumstances be treated with care agents.
5.6

Packaging

Place products into their respective storage devices.
For storage devices, ensure in particular that the covers are correctly and securely seated on the storage
baskets and modules to prevent individual instruments or implants from falling out.
Approved sterilization packaging (e.g. conforming to EN 868, ISO 11607) must be used for sterilization,
subsequent transportation and storage.