Trou ble shooting Guide
Symp toms
1. Unit does not operate
2. Low power output
3. Erratic power output
4. No power output
5. Exces sive leakage current
6. Exces sive power output
7. Voice indi cator announces "Internal fuse."
Power Setting Display flashes "00"
8. Voice indi cator announces "Memory error."
Power Setting Display flashes "01"
9. Voice indicator announces "internal power
error." Power Setting Display flashes "02"
10. Voice indi cator announces "Memory Recycle
error." Power Setting Display flashes "03"
11. Voice indi cator announces "Foot pedal error"
12. Voice indicator announces "internal cut error."
Power Setting Display flashes "04"
13. Voice indicator announces "output error."
Power Setting Display flashes "05"
14. Voice indicator announces "high output error."
Power Setting display flashes "06"
15. Voice indicator announces "internal micro
error." Power Setting Display flashes "07"
16. Voice indi cator announces "Error"
17. Alert LED remains illu mi nated
CODFR08059xx01v05.indd 8
CODFR08059xx01v05.indd 8
Prob able Cause
a. Blown fuse
b. Loose wire at power cord plug
c. No power at wall outlet
a. Low line voltage
b. Incor rect test load
c. Internal cal i bra tion change
a. Loose or dirty con nec tions between forceps
cord and jacks
b. Inter mit tent break in forceps cord
a. Broken wire in forceps cord
a. Shorted output trans former
a. Internal cal i bra tion change
a. Blown internal fuse
a. Oper ating mal func tion
a. Operating malfunction
a. Oper ating mal func tion
a. Two foot pedals pressed simul ta ne ously
a. Operating malfunction
a. Operating malfunction
a. Operating malfunction
a. Operating malfunction
a. Two front panel setting con trols pressed
simulta ne ously
a. Oper ating mal func tion
8
Cor rec tion
a. Replace fuse
b. Check plug for wiring
c. Check electrical service
a. Adjust to nominal line voltage
b. Use 50 ohm non-inductive load for coag and
micro cut; 400 ohm for regular cut
c. Return for service and recal i bra tion
a. Gently clean plug sur faces with abra sive cloth
b. Replace forceps cord
a. Replace forceps cord
a. Return for service and recal i bra tion
a. Return for service and recal i bra tion
a. Replace internal fuse
a. Return for service
a. Return for service
a. Return for service
a. Use care when depressing foot pedal
a. Return for service
a. Return for service
a. Return for service
a. Return for service
a. Use care when pressing con trols
a. Return for service