CARE AND CLEANING
Clean the exterior surface of the 3M™ Attest™ Auto‑reader 490H by wiping it with a
damp cloth moistened with a mild detergent. Do not attempt to clean interior components.
If additional cleaning of the wells is required, call for repair or replacement.
Note: The 3M™ Attest™ Auto‑reader 490H does not contain any user serviceable parts
or require any routine maintenance or calibration. Any malfunctions will be indicated by an
error code. Refer to the Troubleshooting Guide section for error code explanations.
TROUBLESHOOTING GUIDE
In the event of an error code, unplug the 3M™ Attest™ Auto‑reader 490H and plug it back in to initiate the self‑diagnostic test. If error code re‑occurs, call for repair or replacement.
The LCD Display will indicate detected malfunctions by displaying a caution or error code under affected BI wells. The audible alarm, if active, sounds and the LCD Display backlight flashes
to alert the user to all caution (except C1) and error codes. The alarm and backlight can be silenced and extinguished, respectively, by pressing the Buzzer Alarm OFF button
Note that disabling the audible alarm does not disable the flashing LCD Display backlight. Refer to the following:
Code/ Issue
Problem
Unit Warm‑up Caution
C1
Incubation block not at temperature.
BI Removed from Well Caution
C2
3M™ Attest™ BI has been removed before incubation is complete.
C4
Ambient Light Caution
Temperature Control Error
E1
This error will occur if system is no longer able to control the temperature of the
incubator block (60ºC +/‑ 2ºC ).
Memory Error
E3
3M™ Attest™ Auto‑reader 490H unable to save/read data from memory.
Communications Error
E7
This error will occur if internal communications have failed.
LED Error
E8
This error occurs for a specific well if the system detects that a UV LED or Optics
Detection circuit is not working.
BI Inserted Incorrectly Error
E11
Incubation time remaining will not display and fluorescent readings are not taken.
Power Failure Error
E12
Power to Auto‑reader was off for too long to provide valid result for affected BI(s).
Ambient Light Error (occupied well)
E14
Fluorescent result will not display.
Other
Unwanted alarm sounds when positive BI detected.
Alarm Sounds
Other
3M™ Attest™ Biological Indicator has been removed from incubation/reader well
prematurely.
Other
LCD display does not display properly.
Negative Control Vial
Other
BI not properly activated/ media did not contact spore carrier.
8
CAUTION: Always unplug the 3M™ Attest™ Auto‑reader 490H and allow to
cool before cleaning.
Do not spill liquid onto or into the instrument. Do not immerse the unit in liquid.
Clean external surfaces using only the instructions provided by the manufacturer.
Do not open the instrument housing – there are no user serviceable parts. The
instrument must be returned to the manufacturer for repair.
Solution
Allow a 30 minute warm‑up after 3M™ Attest™ Auto‑reader 490H is plugged
in. (Note: 3M™ Attest™ BI readings are disabled until incubator block reaches
60ºC +/‑ 2ºC).
Replace 3M™ Attest™ BI in proper well within 10 seconds to prevent loss of
collected data.
Move 3M™ Attest™ Auto‑reader 490H to reduced ambient light conditions.
Unplug the unit and plug it back in to clear error.
If error doesn't clear, call for repair or replacement.
Unplug the unit and plug it back in to clear error.
If error doesn't clear, call for repair or replacement.
Unplug the unit and plug it back in to clear error. If error doesn't clear, call for
repair or replacement.
Unplug the unit and plug it back in to clear error.
If error doesn't clear, call for repair or replacement. Do not use affected
incubation well.
Remove BI and inspect to verify cap is fully closed, glass ampoule is crushed,
and growth media is in growth chamber. If necessary, repeat BI activation steps
and return BI to incubation well, ensuring it is seated properly.
Ensure power supply connection to Auto‑reader is secure. Retest sterilizer with
a new BI.
Rely on optional visual pH color change result for BI in affected well. Move
3M™ Attest™ Auto‑reader 490H to reduced ambient light conditions to avoid
future occurrences. To clear E14 error code, after visual pH color change result
has been documented and BI has been removed, unplug the 3M™ Attest™
Auto‑reader 490H and plug it back in.
Disable alarm; refer to 3M™ Attest™ Auto-reader 490H Configuration
section for instructions.
Replace 3M™ Attest™ Biological Indicator in incubation/reader well within
10 seconds to prevent loss of data collected.
Unplug the unit and plug it back in. If problem persists call for repair
or replacement.
Ensure BI cap is closed and glass ampoule is crushed. Flick BI until media flows
to spore chamber at bottom of BI.
.