Manufacturer's Information acc. to EN 868-8
Specification of major components
Checks, maintenance and/or
replacements
Container life (sterilization cycles)
Seal life
Cleaning procedures
16
Troubleshooting
Component
Test for
Deformation of the
Excessive load in container
bottom of the
container box
Incorrect loading of the
sterilizer
Excessive pressure
gradients of the sterilizers
Steam inlets are blocked
Discoloration of the
Corrosion of the surface
container box (milky)
Discoloration of the
Corrosion
cover
Gap between cover
Incorrect loading of the
and container box
container
Seals are hard to
Incorrect opening procedure
destroy
used
Residual moisture
Temperature of the
inside the container
sterilization material before
sterilization too low
Sterile container too heavy
Sterilization material stored
incorrectly
Incorrect loading of the
sterilizer
Sterile container used
immediately
Sterile container improperly
stored during cooling down
Insufficient drying effect
Check switching points of
vacuum pump
18
Section 4.1 "Description of the components", page 15 and
Section 2, "Product Overview", page 14
Section 4.2, "Initial Commissioning", page 15
Section 11, "Maintenance and Repair", page 17
Section 13, "Safety Instructions", page 13 and
Section 16, "Troubleshooting", page 18
Acc. to EN 868-8, 4.3
Acc. to EN 868-8, 4.3
Section 5 "Disinfecting, Cleaning and Care", page 16
How/measure
Observe loading height according to section
6, "Container Loading", page 16.
Always store heavy containers at the bottom.
Measure gradients and adjust speed of
pressure change.
Steam inlets (cover and seal area) must not
be covered on the outside.
Use appropriate cleaning agents and
temperatures according to the present
instructions for use.
Do not use acetone, benzine or
glucoprotamin-containing disinfection agents.
Observe loading height according to section 6
"Container Loading", page 16.
Use opening procedure according to the
present instructions for use.
Pre-warm sterilization material to room
temperature (approx. 20°C/68°F).
Observe loading limits as described in section
6.3, "Loading the Container with Instruments",
page 17.
Hollow objects, trays, plates or similar convex
forms should be stored with their openings
facing downward.
Always store heavy containers at the bottom.
Allow container to cool down to room
temperature before.
Do not store container on cold surfaces or in a
draft.
Check drying programs.
Check switching points of vacuum pump.
Check drying time.
Load and unload sterilizer quickly.
marSafe
and marSafe
+ Sterilization container system
®
®
Instructions for Use
Revision 04