Graph
The graph of the calibration is for display purposes only. Values cannot be changed. If the
Correction option is activated, you can modify the correction value for the temperature and/or
moisture using the slide control or by entering numbers.
MW density
The graph shows the deviation of the calibration values from the calculated calibration curve.
Individual calibration points are depicted in accordance with the settings for Type (
with different symbols. Calibration points that are not used for calculating the curve are greyed
out.
O
Note: Calibration measurements can be made whilst the graph is displayed. The input dialogs for
remarks and reference values are displayed automatically.
2. Press on the Calibration-Moisture operation mode or the graphic icon to dis-play the table
instead of the graph.
Zoom
1. Enter a value into the Zoom box or
click the arrow buttons to zoom in or out of the graph.
2. Click the Zoom in or Zoom out control to change the zoom factor by 5% with each click.
3. Once you have zoomed in on the graph, you can move the graphical display by clicking the
graphic and moving it by dragging it with the mouse button held down
4. If no graph is displayed due to the settings, choose auto in Graphic > Settings.
Temp. (correction of temperature)
If a temperature correction has been set for the calibration (
input field labelled Temp. are shown above the graphic.
1. You can now manually change the temperature correction values here in one of the following
ways:
– Moving the slide control
– Manually entering a value
– Pressing the arrow fields
The curves delimiting the correction range are moved on the chart.
98
LMA316PA / LMA321PA - Operator Interface - Reference
Reference density
page 86)
page 100), a slider control and an