Note: If a sensor has several resonance modes, the following rules apply:
Dry product – low frequency
Low density – low frequency
High moisture – high frequency
High density – high frequency
Moisture/Density
Moisture offset/Density offset
An offset is entered if a reference to the calibration of a similar product is used or if the same
difference is always found between the displayed value and the actual (laboratory) value.
O
Example: You enter an offset of 1. If, with the original calibration, a measured value of 10% would
be displayed here, 11% is displayed with the referenced calculation.
Upper/lower limit
Use this entry to specify the limits which trigger an alarm if exceeded by the measured values.
If the box is not selected, measurements outside the range do not trigger an alarm.
MW type
The specifications here serve only to provide information if issues arise. You cannot edit them.
Measurement > Mean
You use this entry to specify the number of results needed to calculate the mean value of each
measuring cycle.
Memory
Specify the amount of memory space to be reserved for results obtained from the product. The
number you enter is approximately equivalent to the number of rows in the table of measured
values.
Select auto if a product is allowed to make complete use of the entire memory available for the
measured values.
If you create several products, share out the memory cells (around 2000) amongst them accord-
ingly. If the set memory capacity is no longer available, the system automatically sets the highest
possible value. If necessary, you can deduct the value required from other products.
You can select the Overwrite option on a product-specific basis. If all memory cells assigned to
this product have been consumed, the oldest values are automatically overwritten (without a
prompt) with the most recent measured values. If the option is not enabled, there will be no further
measuring until you manually delete at least some of the entries.
O
Note: If, when creating a new product, you find that the program only allows you to set a low value
for the memory cells, this is because the products which already exist use up all of the available
memory. If necessary, change the memory settings for the existing products to free memory cells
for the new product.
AB (Linearization/A offset/B offset)
These specifications serve only to provide information if issues arise and should not be edited.
4. Press Apply.
The product has now been created.
O
Note: The number of products which can be created depends on the device version.
LMA316PA / LMA321PA - Operator Interface - Reference