1st display/2nd display
At the top right of the screen, there are two panels for displaying analog readings or statistical
values as required. Settings for what is displayed in these fields can be made independently of
the selection for the main screen.
O
Example: The graphic shows the values for moisture. Temperature and density are shown in the
analog display.
O
Example: The graphic shows the measured values at the cursor at a point on the measurement
curve, whereas the analog display shows the current values.
O
Note: During a calibration run, information on the current calibration is displayed here.
If you wish, you can display statistical values in both of these display panels:
1. Press Analogue display (Details on page 81) or Statistics (Details on page 83) in the menu to
select the type of display.
You can also reach the settings dialog by pressing one of the display panels.
Each display type can be modified if required.
Analogue display
Values are shown as an analog bar chart.
Above the bar, the system tells you whether the given value is the current (online) measurement
or the value at the set cursor position (cursor).
If the current measurement run is stopped and the set display type is Online, the display panel
is set to inactive (greyed out).
O
Note: The cursor position can be changed in the table or graph or by using the controls described
in Navigation (Details on page 72).
The displayed value range is indicated by black marks on each side and the specification of the
value.
The measured value is shown digitally in the middle of the bar chart.
The display is shown in set colours:
Moisture: Blue
Density: Yellow
Temperature: Green
Red markings depict the user-defined alarm range. If the displayed value is in the alarm range,
the bar chart is shown in red.
Settings
You call this dialog from the menu or by pressing the display.
Value
1. Press the list field to select the measured values to be displayed.
2. Choose from the following: