5.2
Connecting the instrument
Connection diagram: Basic configuration (laboratory version)
110-230 V AC
power cable
Attention:When connecting the microwave cables, note the colour markings: Connect the con-
nections marked in green (MW-Gen) on both devices to each other using the correspondingly
marked cable. Use the other cable to connect the connections marked MW-Det.
1. Use the cable marked in green to connect the MW Gen 1 connection of the LMA321PA to the
Gen connection of the sensor.
2. Use the second microwave cable to connect the MW Det 1 connection of the LMA321PA to the
Det connection of the sensor.
3. Link the Sensor 1 connection of both devices using the data cable provided.
4. Connect the supplied keyboard with the PS2 connection marked with the keyboard symbol.
5. Connect the Pt 100 temperature sensor with the socket marked with the thermometer symbol.
Observe the information on page 25.
6. Now establish any other optional connections required for other devices such as a monitor,
mouse, or keyboard (see the overview on the next page).
7. Plug the power cable into the LMA316PA / LMA321PA and then into the power outlet.
24
2x microwave cables
(max. 3 m)
Pt 100 temperature sensor
LMA316PA / LMA321PA - Startup
Sensor
Sensor cable (standard for
Black Line sensors; not required
for older types of sensors)
Keyboard