Sartorius LMA316PA Betriebsanleitung Seite 26

Mikrowellen-feuchtemessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

WLM6009-a10081
Do not use any tools to tighten the connection.
If you want to detach the cable, first complete unscrew the cover from the thread of the device
connection!
5.3
Workflow overview
Once you have connected the evaluation unit, corresponding sensor, and any other peripheral
devices such as a printer, switch the LMA3xxPA / PMD3xxPA on. The software starts up.
You then have to prepare the system for measuring a product as follows:
1. Creating a product (Details on page 30).
2. Calibration (Details on page 35).
Once you have carried out these steps, the device is ready to take measurements:
3. Taking measurements (Details on page 45).
O
Note: Each product to be measured is created once with the corresponding product parameters.
You normally only have to carry out one basic calibration per product and measurement type
(moisture/density). Similar products must be configured separately but can reference a common
calibration.
5.4
Starting the LMA321PA
O
Note: The following description applies to the LMA316PA as well.
1. Turn on the LMA321PA using the on/off switch on the rear of the device.
The software of the LMA321PA microwave measuring system starts up.
O
Note: It may take a couple of minutes for the software to boot. The screen remains black during
this time.
When the user interface appears on the screen the unit is ready to operate.
2. If there is a default user, the software automatically starts up with the rights of that user. You do
not need to log in separately.
If none of the registered users is defined as the default user, the system asks you to enter a user
name and password.
O
Note: To enable you to log in when you start the device for the first time, we will send you a user
name and password separately. We recommend that you then change this password to prevent
unauthorized access (see page 49).
Recommendation: After switching on the device, wait a few minutes before taking the first meas-
urements, since this allows the device to reach its optimum operating temperature.
5.5
Turning the LMA321PA off
When you have finished your work, simply turn off the LMA321PA using the on/off switch on the
rear of the device. You do not need to close the operating software first.
To restart the operating software without turning off the device, press Service > Reset. Note that
you need at least Level 4 access rights to do this.
26