Note: The currently selected product is indicated by a green tick in the No. column.
1. Press the column header to sort the table. The current sort order is indicated by an arrow in the
column header.
2. Press the row containing the name of the product to be measured.
O
Note: Products with similar properties can reference the same calibration. If a product references
the calibration of a similar product, corresponding information can be found in the Reference to
calibration line. In the case of products with their own calibration, the Cal. link column contains
the entry self.
O
Note: Press the navigation controls (see Navigation (Details on page 72)) to view products you
cannot see immediately.
3. Press Apply. The measurement now takes place for the selected product.
Table
The measurement values are displayed in a table.
1. Press the column header to sort the table. The current sort order is indicated by an arrow in the
column header.
O
Note: The first column (without title) shows the row number (position) dependent on the sorting.
The No. column shows the automatically allocated sequential number for the measurement
value.
O
Note: You can customize the appearance of the table by choosing Measurement > Table >
Settings (Details on page 76). Here, you can determine whether the table can be sorted or
whether the columns can be moved.
2. Use the controls described in Navigation (Details on page 72) to show further values in the
measurement series.
3. Press the Measurement mode display or the table symbol to switch to the graphical display.
4. To start or stop the measurement process and to take samples and carry out an empty sensor
adjustment, use the controls described in Measurement/Calibration (Details on page 70).
Determining the cursor position
5. Press a row. The row is highlighted by a red border. When you switch to the graphical display
(
page 77), the cursor position is retained.
O
Note: Press the Goto last control once.
The control remains activated, and therefore has a blue background. The most recently
measured value is displayed in the bottom row. The table is constantly refreshed.