Individual values can be deleted from the calibration.
In the menu, select Calibration > Density calibration > Table > Edit > Delete.
The table is now displayed in delete mode, i.e. the controls (
rows are displayed at the bottom.
1. Press a row. The value is selected.
2. Press Delete.
The selected value is deleted. The system continues to take the value into account until the cal-
ibration curve is recalculated.
3. Press Close.
The deletion process is complete.
O
Note: Calibration points can be removed from the calibration without being deleted. To do so,
choose Type and set the point's property to Invalid (
account for the calibration.
O
Note: If you have a previous calibration that you are satisfied with, you can delete the individual
values to free up memory space. The actual calibration is retained.
Copy
Individual calibration values can be copied.
In the menu, select Calibration > Density calibration > Table > Edit > Copy.
The table is now displayed in copy mode, i.e. the controls (
rows are displayed at the bottom.
Copied entries are pasted to the calibration data of a product or to the settings for the density test
as selected.
1. Select an entry, a block, or all entries.
To select several rows, press Multi.
Now select several values using the Block 1/Block 2 control or select entries individually.
2. Press Copy.
3. Select a product from the Copy to dialog.
4. Specify whether the selected values are to be copied to the calibration of the product or the com-
pression test.
Only the measured values are copied - not the reference values.
5. Press Close.
The copy process is complete.
New
The Calibration properties dialog appears.
You create a new entry in the calibration table independently of the measured values. Every value
can be entered via this dialog.
1. Press an input field to display the Numerical input (Details on page 125) or Text input (Details
on page 124) dialog.
2. Press Apply to insert the entry into the table.