Compression test properties
1. The Ranges tab allows you to set the parameters for the measurement ranges in which readings
are taken (displayed in green in the graph) and for the measurement range indicating the empty
sensor state (displayed in red in the graph).
2. On the Scale/Graph Type/Regression tab, you define parameters for modifying the depiction of
the graphic.
Select free for Regression and press Apply: The measuring points are connected by a line.
The readings found should all be positioned on a a straight line going through the neutral point,
if possible. Please contact the Sartorius customer service if this cannot be achieved.
3. Once you are satisfied with the results, press the Submit to sensor mode control.
The sensor mode you selected is adapted to the test results.
Print values
Selected measurement values can be printed out. See page 53 for information on how to connect
a printer.
Choose Service > MW test > Compression test > Table > Print.
The table is now displayed in print mode, i.e. controls for selecting one or more rows are dis-
played at the bottom.
1. Select (highlight) the lines you wish to print.
2. Press the Print control.
The selected rows are printed out.
3. Press Close. The printing of the values is complete.
34
LMA316PA / LMA321PA - Calibration