The dialog for entering reference values now appears. How you proceed hereafter depends on
whether you used samples with existing reference values or whether the samples are now being
sent to your laboratory for the determination of the reference values.
Option 1: Reference values exist
14. Enter the reference value for the sample and press the Apply button (green tick). The measured
value is displayed in the table.
O
Note: Even if several measurements are taken for a sample, only one line with the average value
of the measurements is shown.
Option 2: Reference values do not yet exist
15. Leave the input field in the Reference moisture or Reference density dialog empty and press
Cancel. The measured value is displayed in the table without reference values.
16. Send the samples for which you have measured calibration values to the laboratory for the deter-
mination of the reference moisture or density.
O
Note: The accuracy of the calibration and thus of all later measurements depends on the
accuracy of the reference values. The moisture level of the samples must not change during
transport and measurement in the laboratory. Observe the information on calibration (from page
29).
As soon as you obtain the reference values for the measured samples, you enter them into the
calibration table as follows:
17. In the menu, select Calibration > Moisture Calibration or Density calibration > Table > Edit
> Values.
18. In the submenu, select either Reference moisture or Reference density.
O
Note: Alternatively, you can choose All values. When you press the Edit control (see below), the
Calibration properties dialog appears and you can edit all values for this calibration.
The table is now displayed in edit mode.
19. Press the relevant row to select it.
20. Press the control to start editing.
The Numerical input (Details on page 125) or Text input (Details on page 124) dialog appears.
21. Enter a value.
22. Press the Apply control (green tick).
LMA316PA / LMA321PA - Calibration